MF010, 020 Series Datasheet by Honeywell Sensing and Productivity Solutions | Digi-Key Electronics

MF010, 020 Series Datasheet by Honeywell Sensing and Productivity Solutions

View All Related Products | Download PDF Datasheet
Honeywell Temperature
MF010 Series Low
Temperature and
MF020 Series
High
Temperature
Oxygen Sensors
DESCRIPTION
The MF010 Series Low Temperature Oxygen Sensors is
designed to determine the oxygen concentration in air or inert
gas mixtures with temperatures of -100 °C to 250 °C [-148 °F
to 482 °F] max. The MF020 Series High Temperature Oxygen
Sensors is designed to determine the oxygen concentration in
air or inert gas mixtures with temperatures of -100 °C to 400 °C
[-148 °F to 752 °F] max.
These products are particularly suitable for measuring oxygen
in areas that are not easily accessible, or in closed systems.
The MF010 and MF020 Series may both be supplied with a
measuring range of 0.1 vol%O2 to 25 vol%O2 and 0.1 vol%O2
to 100 vol%O2. The entire measurement range is linear for
both series.
The sensor detects the gas mixtures oxygen partial pressure
(absolute oxygen content) by using a dynamic process that
occurs on two zirconium dioxide discs which form a
hermetically sealed chamber. Because no chemical
substances are used, the sensor exhibits significantly longer
service life than an electrochemical sensor.
The actual oxygen sensor is mounted in the tip of the bar
probe and is protected by a stainless-steel sintered disk which
serves as a flame back-flash stop. The water-proof plastic
housing accommodates the electronics and is mechanically
connected to the bar probe.
Sensor electronics include:
1. Signal amplifier
2. Control for the ionic pump with an analog component
3. Test component
4. Internal monitoring logic component
5. Power supply for the sensor heating element and the
analog and digital components
6. Reset
7. Voltage monitor
8. Analog output
9. Bi-directional digital output
10. Internal potentiometer
The sensor outputs the measured values via an analog
channel (4 mA to 20 mA or 0 V to 10 V) and a digital channel
(also provides any error signals) which are evaluated and
further processed in a user-specified external mounting device
such as a fuel controller, display, measuring instrument,
programmable logic controller or ventilation system.
Both series may be calibrated manually or self-calibrated using
atmospheric air instead of a reference gas. The sensor can
also conduct external operational self-tests.
Both series may also operate in a fault-proof mode. The sensor
self-monitors during operation and provides error signal output.
A second oxygen sensor is not necessary.
FEATURES
No moving parts
Low power consumption
No need for temperature stabilization
Functional testing and calibration in ambient air
Long life
POTENTIAL APPLICATIONS
Boiler combustion controls
Aircraft oxygen generation systems
MFO10-2-LC1 MFO10-2—LCZ MFO10-2-LCZ MF010-2-LC4 MFOZO-Z-LC1 MFOZO-Z-LCZ MFOZO-Z-LCZ MF020-2-LC4
MF010 Series and MF020 Series
2 www.honeywell.com/sensing
Table 1. MF010 Series Specifications and Order Information
Characteristic
Catalog Listing
MF010-0-LC1
MF010-1-LC1
MF010-2-LC1
MF010-1-LC2
MF010-0-LC3
MF010-1-LC3
MF010-2-LC3
MF010-0-LC4
MF010-1-LC4
MF010-2-LC4
Temperature range at probe tip
-100 °C to 250 °C [-148 °F to 482 °F]
Output
0 Vdc to 10 Vdc
4 mA to 20 mA
O
2
range (O
2
partial pressure)
with equivalent O2 output
1 mbar to 250 mbar
0.1 vol%O2 to 25 vol%O2
1 mbar to 250 mbar
0.1 vol%O2 to 25 vol%O2
1 mbar to 1000 mbar
0.1 vol%O2 to 100 vol%O2
Probe length:
-
0-LXX
-
1-LXX
-2-LXX
220 mm [8.7 in]
400 mm [15.7 in]
600 mm [23.6 in]
Table 2. MF020 Series Specifications and Order Information
Characteristic
Catalog Listing
MF020-0-LC1
MF020-1-LC1
MF020-2-LC1
MF020-1-LC2
MF020-0-LC3
MF020-1-LC3
MF020-2-LC3
MF020-0-LC4
MF020-1-LC4
MF020-2-LC4
Temperature range at probe tip
-100 °C to 400 °C [-148 °F to 752 °F]
Output
0 Vdc to 10 Vdc
4 mA to 20 mA
O
2
range (O
2
partial pressure)
with equivalent O2 output
1 mbar to 250 mbar
0.1 vol%O2 to 25 vol%O2
1 mbar to 250 mbar
0.1 vol%O2 to 25 vol%O2
1 mbar to 1000 mbar
0.1 vol%O2 to 100 vol%O2
Probe length:
-
0-LXX
-
1-LXX
-2-LXX
220 mm [8.7 in]
400 mm [15.7 in]
600 mm [23.6 in]
Table 3. MF010 and MF020 Series General Specifications
Characteristic
Parameter
Supply voltage
24 Vdc ±5%
Supply current
500 mA
Ambient temperature (includes solar radiation
-10° C to 50° C [14 °F to 122 °F]
Resolution
12 bit
Heat-up time
approx. 10 min
Accuracy
±2% FS (Full Scale)
Reproducibility
±1% FS (Full Scale)
Gas flow rate
0 m/s to 10 m/s
Reaction time
approx. 3 s
Incidental permissible acceleration
30 g
Repetitive permissible acceleration
5 g
Bar probe material
stainless steel 1.4301
Housing material
Makrolon 8030 (30% GV), UL94 V-1
Housing sealing
IP65
Connector sealing
IP67
Weight (without bar probe)
150 g [5.3 oz]
Termination (female connector)
99-4226-14-07 (Binder 693 series)
Certification
complies with EMC directives EN 50082-2 and EN 50081-1 and directives
89/336/EWG and 92/31/EWG
Honeywell :7? ””””” i l __>7 H7 H a ”Saab $39577 +W@+
Oxygen Sensors
Honeywell Sensing and Control 3
Figure 1. Dimensional Drawing and Pinout (For reference only: mm/in.)
21,0
[0.83]
27,7
[1.09]
88,0
[3.46]
105,0
[4.13]
60,0
[2.36]
30,0
[1.18]
21,0
[0.83] 42,0
[1.65]
Ø12,0
[0.47] Ø9,0
[0.35] Ø0,21
[5.3]
Ø26,5
[1.04]
35,5
[1.38]
17,5
[0.69]
42,5
[1.67]
14,0
[0.55]
Varies with catalog listing.
See Table 1.)
Pin
+24 Vdc
ground
0V
not assigned
Channel 1
(4 mA to 20 mA)
1
2
3
4
5
6
7
Channel 2 (I/O)
input +24 Vdc
6
5
1
72
43
5V
0V
0Vor4mA=0Vol%O
2
10Vor20mA=25Vol%O
2
calibration switch
Figure 2. System Block Diagram
Figure 3. Sensor Mounting
MF010/MF020
Sensor
Power
Supply
24 Vdc
User-Specified
External Monitoring
Device
(4 mA to 20 mA
or 0 V to 10 V
Output)
1 kOhm + LED
Ground
Test
0V
Calibration Connection Cable
Partition
Housing
Bar Probe
Gas Mixture
250 °C [482 °F] Max. (MF010)
400 °C [752 °F] Max. (MF020)
Atmospheric Air
50 °C [122 °F] Max.
15°
Weld Socket
O-Ring
Nut
Bar Probe
Partition
Order Guide
Catalog Listing
Description
MF010 Series
Low temperature oxygen sensor (See Table 1 for specifics.)
MF020 Series
High temperature oxygen sensor (See Table 2 for specifics.)
Honeywell
Sensing and Control
Honeywell
1985 Douglas Drive
North
Minneapolis, MN 55422
www.honeywell.com/sensing
000670
-3-EN IL50 GLO Printed in USA
July
2011
Copyright © 2011 Honeywell International Inc. All rights reserved.
WARNING
PERSONAL INJURY
DO NOT USE these products as safety or emergency stop
devices or in any other application where failure of the
product coul
d result in personal injury.
Failure to comply with these instructions could result in
death or serious injury.
WARRANTY/REMEDY
Honeywell warrants goods of its manufacture as being free of
defective materials and faulty workmanship. Honeywell’s
standard product warranty applies unless agreed to otherwise
by Honeywell in writing; please refer to your order
acknowledgement or consult your local sales office for specific
warranty details. If warranted goods are returned to Honeywell
during the period of coverage, Honeywell will repair or replace,
at its option, without charge those items it finds defective. The
foregoing is buyer’s sole remedy and is in lieu of all other
warranties, expressed or implied, including those of
merchantability and fitness for a particular purpose. In no
event shall Honeywell be liable for consequential, special,
or indirect damages.
While we provide application assistance personally, through
our literature and the Honeywell web site, it is up to the
customer to determine the suitability of the product in the
application.
Specifications may change without notice. The information we
supply is believed to be accurate and reliable as of this
printing. However, we assume no responsibility for its use.
WARNING
MISUSE OF DOCUMENTATION
The information presented in this product sheet is for
reference only. Do not use this document as a product
installation guide.
Complete installation, operation, and maintenance
information is provided in the instructions supplied with
each product.
Failure to comply with these instructions could result in
death or serious injury.
SALES AND SERVICE
Honeywell serves its customers through a worldwide network
of sales offices, representatives and distributors. For
application assistance, current specifications, pricing or name
of the nearest Authorized Distributor, contact your local sales
office or:
E-mail: info.sc@honeywell.com
Internet: www.honeywell.com/sensing
Phone and Fax:
Asia Pacific +65 6355-2828
+65 6445-3033 Fax
Europe +44 (0) 1698 481481
+44 (0) 1698 481676 Fax
Latin America +1-305-805-8188
+1-305-883-8257 Fax
USA/Canada +1-800-537-6945
+1-815-235-6847
+1-815-235-6545 Fax

Products related to this Datasheet

SENSOR OXYGEN 24V 500MA
SENSOR OXYGEN 24V 500MA
SENSOR OXYGEN 24V 500MA
SENSOR OXYGEN 24V 500MA
MF010 LOW TEMP OXYGEN SENSOR
MF010 LOW TEMP OXYGEN SENSOR
MF010 LOW TEMP OXYGEN SENSOR
MF010 LOW TEMP OXYGEN SENSOR
MF010 LOW TEMP OXYGEN SENSOR
MF010 LOW TEMP OXYGEN SENSOR
MF010 LOW TEMP OXYGEN SENSOR
MF010 HIGH TEMP OXYGEN SENSOR
MF010 HIGH TEMP OXYGEN SENSOR
MF010 HIGH TEMP OXYGEN SENSOR
MF010 HIGH TEMP OXYGEN SENSOR
MF010 HIGH TEMP OXYGEN SENSOR
MF010 HIGH TEMP OXYGEN SENSOR
MF010 HIGH TEMP OXYGEN SENSOR
MF010 HIGH TEMP OXYGEN SENSOR
MF010 HIGH TEMP OXYGEN SENSOR
MF010 HIGH TEMP OXYGEN SENSOR
MF010 HIGH TEMP OXYGEN SENSOR
MF010 HIGH TEMP OXYGEN SENSOR
SENSOR GAUGE
SENSOR OXYGEN
SENSOR OXYGEN